Bangladesh BEST Programme Calibration of Micro-pipette By

17 Slides142.29 KB

Bangladesh BEST Programme Calibration of Micro-pipette By Gravimetric Method Nihal Gunasekara Sri Lanka

Bangladesh BEST Programme Types of Micro-pipettes Single Channel Multi- Channel

Bangladesh BEST Programme Requirements Unit of volume Cubic millimeter : µl Ref. temperature 20 C To deliver Apparatus is calibrated to accurately transfer the stated volume to another container.

Bangladesh BEST Programme Apparatus Needed Balance Thermometer Hygrometer Barometer Distilled or deionized water Weighing vessel

Bangladesh BEST Programme Recommended Apparatus - Balance Selected Volume (V) 1 µl V 10 µl 10 µl V 100 µl Resolution of Balance (mg) Std. Deviation (mg) 100 µl V 1000 µl 0.001 0.01 0.1 0.002 0.02 0.2 1 ml V 10 ml 0.1 0.2

Bangladeshi BEST Programme Recommended Apparatus/Liquids Apparatus/Liquid Thermometer Hygrometer Barometer Calibration Liquid Uncertainty 0.2 C 10% 0.5 kPa Distilled /de-ionized water

Bangladesh BEST Programme Conditions for the Test Temperature Maintained at 20 0.5 C and during M calibration 0. 2 C stability Humidity Maintained 50 % Prefer to maintain at 70 10 % Stabilization Volumetric instruments and test water 1 to 2 hours Evaporation Special precautions below 50 µl Test cycle Less than 60 s

Bangladesh BEST Programme Test Volume Fixed Volume M Nominal Volume Variable Volume At least three volumes Preferably 10%, 50 % and nominal volume Measurements 10 at each volume

Bangladesh BEST Programme Calibration Procedure- Single Channel Place the test liquid in the weighing vessel to a depth of at least up to 3mm and record the temperature, barometric pressure & humidity Fit the selected disposable tip to the micro pipette barrel Fill the tip with test liquid and expel to waste five time to reach a humidity equilibrium in the dead air volume Place the weighing vessel with its added water on the balance pan

Bangladesh BEST Programme Calibration Procedure- Single Channel Prior to any calibration test make sure that all instruments reached the ambient temperature in the laboratory During the calibration maximum temperature variation in the laboratory should not exceed 0.2oC. Be sure that all the apparatus use for calibration and the distilled water are at room temperature.

Bangladesh BEST Programme Calibration Procedure Replace the disposable tip of the piston pipette Fill the piston pipette with test liquid, immersing its delivery orifice 2 to 3 mm below the surface of water Release the operating button slowly and draw the pipette vertically against the side wall of container . Expel the water to waste to pre-wet the tip and refill the piston pipette as above. Record the mass m0 of the weighing vessel or tare the balance to zero Deliver the contents of the pipette into the weighing vessel by touching pipette tip against the container wall just above liquid surface at an angle 30 to 45 Record the mass m1 of the weighing vessel and determine the quantity delivered Repeat the above cycle for ten measurements

Bangladesh BEST Programme Calibration Procedure- Multi-Channel Procedure is similar to single channel micro pipette During the calibration each channel should be considered as a single channel

Bangladesh BEST Programme Conversion of Mass into Volume vi mi Z The Z correction factor is a function of test temperature and air pressure 13

Bangladesh BEST Programme Z Correction Factors for Distilled Water (Z values in micro-litres per milligram ) Temp. Air Pressure kPa C 80 85 90 95 100 101.3 105 17 17.5 1.0021 1.0022 1.0021 1.0022 1.0022 1.0023 1.0022 1.0023 1.0023 1.0024 1.0023 1.0024 1.0023 1.0024 18.0 18.5 1.0022 1.0023 1.0023 1.0024 1.0023 1.0024 1.0024 1.0025 1.0025 1.0025 1.0025 1.0026 1.0025 1.0026 19.0 19.5 1.0024 1.0025 1.0025 1.0026 1.0025 1.0026 1.0026 1.0027 1.0026 1.0027 1.0027 1.0028 1.0027 1.0028 20.0 20.5 1.0026 1.0027 1.0027 1.0028 1.0027 1.0028 1.0028 1.0029 1.0028 1.0029 1.0029 1.0030 1.0029 1.0030 21.0 21.5 1.0028 1.0030 1.0029 1,0030 1.0029 1.0031 1.0030 1.0031 1.0031 1.0032 1.0031 1.0032 1.0031 1.0032 22.0 22.5 1.0031 1.0032 1.0031 1.0032 1.0032 1.0033 1.0032 1.0033 1.0033 1.0034 1.0033 1.0034 1.0033 1.0034

Bangladesh BEST Programme Mean Volume 1 10 V Vi 10 i 1 15

Bangladesh BEST Programme Data Recording Use spread sheet for data recording Record Temperature, Humidity, Barometric pressure Record final data in a data recording sheet Review from a third party Prepare the final calibration certificate

Bangladesh BEST Programme 17