Oxygen Delivery Devices they’re not what you think

19 Slides86.50 KB

Oxygen Delivery Devices they’re not what you think

Indications for oxygen use Low ambient O2 Hypoxemia Suspected hypoxemia Increased work-of-breathing Flaring Tachypnea Increased myocardial work Acute head injury

O2 delivery systems High-flow Venti-masks (air entrainment masks) Mechanical aerosol systems High-flow humidifier systems Low-flow Nasal cannula Simple masks Partial rebreathing masks Non-rebreathing masks

Entrainment ratios Entrainment ratio FiO2 Minimal flow to operate Total flow (L/min) 1/25 0.24 4 104 1/10 1/7 1/5 0.28 0.31 0.35 4 6 8 44 48 48 1/3 0.40 8 32 1/1.7 0.50 12 32 1/1 1/0.6 0.60 0.70 12 12 24 19

Delivered FiO2 with low-flow

What’s my PaO2? a: b: On room air? On 0.50 FiO2? c: On 0.80 FiO2?

The answers: PaO2 FiO2 x (barometric pressure water vapor pressure) - PaCO2/RQ PaO2 FiO2 X (760 - 43) - PaCO2/0.8 a: PaO2 .21 X 713 - 40/0.8 100 b: PaO2 .50 X 713 - 40/0.8 306 c: PaO2 .80 X 713 - 40/0.8 520

O2Hgb dissociation curve 100 % Sat 80 60 40 20 0 0 20 40 60 80 100 PO2 mmHg 120 140 160

Resorption atelectasis 100% O2 nitrogen oxygen A A B A B PO2 673 PCO2 40 PH2O 47 B After 15 minutes, blood N2 is depleted. Poorly ventilated & well perfused units (A) become atelectactic.

Non-rebreathing system Room air Gas source Expiratory gas Reservoir To patient One way valves

Air entrainment devices exhaled gas oxygen room air

O2Hgb dissociation curve shift 100 % Sat 80 pH CO2 60 temp 40 20 0 0 20 40 60 80 100 PO2 mmHg 120 140 160

What happens to your pO2 and sats when your pH changes? ph 7.4, pO2 100, sat 96% ph 7.0, pO2 ?, sat ?

Oxygen delivery # trucks X how full they are loaded X how fast they go DO2 1.38(hgb X sat X CI) (pO2 X 0.003)

Entrainment ratios System Ventimasks Aerosol Flow Total Entrainment rate flow FiO2 ratio (O2/air) (L/min) (L/min) 0.24 1-25 4 104 0.28 1-10 4 44 0.31 1-7 6 48 0.35 1-5 8 48 0.4 1-3 8 32 0.5 1-1.7 12 32 0.6 1-1 12 24 0.7 1-0.6 12 19

Estimating FiO2 O2 Flow rate FiO2 Nasal cannula 1 0.24 2 0.28 3 0.32 4 0.36 5 0.4 6 0.44 O2 Flow rate FiO2 Oxygen mask 5-6 0.4 6-7 0.5 7-8 0.6 O2 Flow rate FiO2 Mask with reservoir 6 0.6 7 0.7 8 0.8 9 0.80 10 0.80

Simple mask Exhalation ports O2 inlet Open ports for exhaled gas Air entrained through ports if O2 flow through does not meet peak insp flow 35-55% O2 at 6-10 L/min

Partial rebreather mask Exhalation ports O2 Reservoir O2 directed into reservoir Insp: draw gas from bag & ? room air Exp: first 1/3 of exhaled gas goes into bag (dead space) Dead space gas mixes with ‘new’ O2 going into bag Deliver 60% O2

Non-rebreathing mask One-way valves O2 Reservoir Valve prevents exhaled gas flow into reservoir bag Valve over exhalation ports prevents air entrainment Delivers 100% O2, if bag does not completely collapse during inhalation